Diesel #2/Mineral Oil Standard on Rxi-5HT (15 m x 0.25 mm x 0.10 µm)

GC_EV1378
PeakstR (sec)
1.C12203.5
2.C16349.2
3.C20468.8
ColumnRxi-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905)
Standard/SampleDiesel #2/mineral oil (cat.# 31676)
Diluent:Hexane
Conc.:5,000 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Split Vent Flow Rate:17.5 mL/min
Oven
Oven Temp.:40 °C (hold 0.1 min) to 400 °C at 19.6 °C/min (hold 1.53 min)
Carrier GasH2, constant flow
Flow Rate:1.75 mL/min
DetectorFID @ 420 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesActual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc).