Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Detailed Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) in Creosote Oil on Rxi-PAH

ColumnRxi-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
SampleCreosote oil standard (cat.# 31838)
Diluent:Dichloromethane
Conc.:50,000 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 40:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Split Vent Flow Rate:78 mL/min
Oven
Oven Temp.:40 °C (hold 1.6 min) to 210 °C at 24 °C/min to 295 °C at 1.9 °C/min to 350 °C at 3.7 °C/min (hold 6 min)
Carrier GasHe, constant flow
Flow Rate:1.95 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
13.435-4503.5
22535-4501.8
Transfer Line Temp.:330 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9mm ID
Source Temp.:350 °C
Quad Temp.:200 °C
Electron Energy:70 eV
Solvent Delay Time:3.4 min
InstrumentAgilent 7890B GC & 5977A MSD