Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

NIST SRM 1975, Diesel Particulate Extract, on Rxi-PAH

ColumnRxi-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
Standard/SampleNIST SRM 1975
Diluent:Dichloromethane
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Split Vent Flow Rate:19.5 mL/min
Oven
Oven Temp.:110 °C (hold 1.6 min) to 210 °C at 24 °C/min to 295 °C at 1.9 °C/min to 350 °C at 3.7 °C/min (hold 6 min)
Carrier GasHe, constant flow
Flow Rate:1.95 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
13.09102.1, 108.1, 128.1, 136.2 40
25.69115.1, 142.1 40
36.2376.1, 141.1, 154.1, 156.2 40
46.9575.6, 76.1, 91.1, 152.1, 153.1, 155.1, 162.2, 170.2 20
57.9382.4, 165.1 40
69.19139.1, 152.1, 160.2, 178.1, 184.1, 188.2 40
711.7894.6, 165.1, 190.1, 192.1 40
814.19101.1, 202.1 40
916.60101.1, 202.1 40
1018.4992.1, 184.1 40
1119.58108, 216 40
1222.3196.1, 212.2 40
1324.92113.1, 226.1, 228.1 40
1426.27113.1, 114.0, 120.1, 226.1, 228.1, 240.1 40
1528.75154.1, 252.1 35
1630.62119.8, 242.2 40
1735.29126.1, 252.1 40
1838.56125.1, 252.1 40
1939.23126.1, 252.1 40
2040.71125.1, 126.1, 132.1, 252.1, 264.1 40
2142.91125, 132.2, 252.1, 264.1 40
2244.35252.1, 268.1 40
2348.41139.1, 139.5, 278.1, 279.1 40
2452.13139.1, 139.5, 278.1, 279.1 40
2553.70138.1, 139.1, 278.1 40
2654.97138.1, 276.1 40
2755.92138.1, 276.1 40
2857.04132.6, 267.1 40
2959.69151.0, 302.1 40
3063.27150, 151, 156.1, 300.1, 302.1, 312.1 40
3164.78151.0, 302.1 40
Transfer Line Temp.:330 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9 mm ID
Source Temp.:350 °C
Quad Temp.:200 °C
Solvent Delay Time:3 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD