Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

EPA and EU Phthalates on Rxi-XLB

ColumnRxi-XLB, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13723)
SampleBenzyl benzoate (cat.# 31847)
EPA Method 8061A phthalate esters mixture (cat.# 33227)
Diluent:Methylene chloride
Conc.:50.0 µg/mL (80 μg/mL for internal standard benzyl benzoate)
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 3.5 mm Precision liner w/wool (cat.# 23320.1)
Inj. Temp.:280 °C
Split Vent Flow Rate:3 mL/min
Oven
Oven Temp.:200 °C (hold 0.5 min) to 330 °C at 30 °C/min (hold 1 min)
Carrier GasHe, constant linear velocity
Linear Velocity:66.7 cm/sec, 39.5 psi, 272.3 kPa @ 200 °C
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
10.959-400-
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Temp.:280 °C
Electron Energy:70 eV
Solvent Delay Time:0.9 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
NotesThe flow rate is 3 mL/min @ 200 °C. The MS scan interval is 0.1 sec.
AcknowledgementThe authors would like to thank Shimadzu Corporation for their consultation with the operation of the QP2010 Plus GC-MS instrument.