Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

EPA Method 530 UCMR4 Standard at 10x the Method Reporting Limit on Rtx-1701 (SIM)

GC_EV1467
PeakstR (min)Conc.
(µg/mL)
1.o-Toluidine-d9 (SS)8.431.0
2.o-Toluidine8.480.70
3.Quinoline-d7 (SS)10.621.0
4.Quinoline10.642.0
5.Acenaphthene-d10 (IS)13.451.0
6.Butylated hydroxyanisole (BHA)14.253.0
7.Phenanthrene-d10 (IS)17.321.0
ColumnRtx-1701, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12023)
SampleMethod 530 UCMR4 standard (cat.# 572262)
Method 530 UCMR4 surrogate standard (cat.# 572265)
Method 530 UCMR4 internal standard (cat.# 572266)
Diluent:Dichloromethane
Injection
Inj. Vol.:1 µL pulsed splitless (hold 1.0 min)
Liner:Topaz 4 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:275 °C
Pulse Pressure:20 psi (137.9kPa)
Pulse Time:1.05 min
Purge Flow:60 mL/min
Oven
Oven Temp.:60 °C (hold 1 min) to 290 °C at 10 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
13.50106, 107, 112, 114 25
29.55102, 108, 129, 136 25
312.08162, 164 25
414.02137, 180 25
515.92160, 188 25
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Stainless Steel
Drawout Plate:6 mm ID
Source Temp.:280 °C
Quad Temp.:180 °C
Solvent Delay Time:1.45 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentHP6890 GC & 5973 MSD
NotesThe EPA Method 530 UCMR4 standard analyte concentrations vary to simplify preparing ICAL levels based on the minimum method reporting levels.