Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Semivolatiles on Rxi-5Sil MS by U.S. EPA Method 8270 Using the GC Accelerator Kit and Split Injection with a 208 V GC Oven

GC_EV1475
PeakstR (min)
1.1,4-Dioxane-d8 (IS)1.32
2.N-Nitrosodimethylamine1.43
3.Pyridine1.45
4.2-Fluorophenol (SS)1.87
5.Phenol-d6 (SS)2.20
6.Phenol2.20
7.Aniline2.23
8.Bis(2-chloroethyl) ether2.24
9.2-Chlorophenol2.28
10.1,3-Dichlorobenzene2.34
11.1,4-Dichlorobenzene-d4 (IS)2.35
12.1,4-Dichlorobenzene2.36
13.Benzyl alcohol2.40
14.1,2-Dichlorobenzene2.42
15.2-Methylphenol2.44
16.Bis(2-chloroisopropyl)ether2.45
17.4-Methylphenol2.50
18.3-Methylphenol2.50
19.N-Nitrosodi-N-propylamine2.50
20.Hexachloroethane2.56
21.Nitrobenzene-d5 (SS)2.57
22.Nitrobenzene2.58
23.Isophorone2.67
24.2-Nitrophenol2.71
25.2,4-Dimethylphenol2.71
26.Benzoic acid2.74
27.Bis(2-chloroethoxy)methane2.75
28.2,4-Dichlorophenol2.80
29.1,2,4-Trichlorobenzene2.84
30.Naphthalene-D8 (IS)2.87
31.Naphthalene2.88
PeakstR (min)
32.4-Chloroaniline2.89
33.Hexachlorobutadiene2.93
34.4-Chloro-3-methylphenol3.09
35.2-Methylnaphthalene3.17
36.1-Methylnaphthalene3.21
37.Hexachlorocyclopentadiene3.23
38.2,4,6-Trichlorophenol3.28
39.2,4,5-Trichlorophenol3.30
40.2-Fluorobiphenyl (SS)3.32
41.2-Chloronaphthalene3.38
42.2-Nitroaniline3.41
43.1,4-Dinitrobenzene3.46
44.Dimethyl phthalate3.49
45.1,3-Dinitrobenzene3.50
46.2,6-Dinitrotoluene3.51
47.1,2-Dinitrobenzene3.54
48.Acenaphthylene3.56
49.3-Nitroaniline3.59
50.Acenaphthene-d10 (IS)3.62
51.Acenaphthene3.63
52.2,4-Dinitrophenol3.63
53.4-Nitrophenol3.65
54.2,4-Dinitrotoluene3.69
55.Dibenzofuran3.70
56.2,3,5,6-Tetrachlorophenol3.73
57.2,3,4,6-Tetrachlorophenol3.75
58.Diethyl Phthalate3.79
59.4-Chlorophenyl phenyl ether3.84
60.Fluorene3.85
61.4-Nitroaniline3.85
62.4,6-Dinitro-2-methylphenol3.86
PeakstR (min)
63.N-Nitrosodiphenylamine (as diphenylamine)3.89
64.N-N-Diphenylhydrazine (as azobenzene)3.91
65.2,4,6-Tribromophenol (SS)3.95
66.4-Bromophenyl phenyl ether4.06
67.Hexachlorobenzene4.09
68.Pentachlorophenol4.17
69.Phenanthrene-d10 (IS)4.26
70.Phenanthrene4.27
71.Anthracene4.29
72.Carbazole4.36
73.di-n-Butyl phthalate4.49
74.Fluoranthene4.79
75.Pyrene4.79
76.Benzidine4.84
77.Pyrene-d10 (IS)4.89
78.p-Terphenyl-d14 (SS)4.96
79.Butyl benzyl phthalate5.19
80.3,3'-Dimethylbenzidine5.20
81.Bis(2-ethylhexyl) adipate5.22
82.3,3'-Dichlorobenzidine5.52
83.Bis(2-ethylhexyl) phthalate5.53
84.Benz[a]anthracene5.55
85.Chrysene-d12 (IS)5.56
86.Chrysene5.58
87.Di-n-octyl phthalate6.02
88.Benzo[b]fluoranthene6.40
89.Benzo[k]fluoranthene6.42
90.Benzo[a]pyrene6.71
91.Perylene-d12 (IS)6.77
92.Indeno[1,2,3-cd]pyrene7.98
93.Dibenz[a,h]anthracene8.01
94.Benzo[ghi]perylene8.30
ColumnRxi-5Sil MS, 20 m, 0.15 mm ID, 0.15 µm (cat.# 43816)
Sample8270 MegaMix (cat.# 31850)
8270 Benzidines mix (cat.# 31852)
Benzoic acid (cat.# 31879)
Revised B/N surrogate mix (cat.# 31888)
Acid surrogate mix (4/89 SOW) (cat.# 31063)
Revised SV internal standard mix (cat.# 31886)
Diluent:Methylene chloride
Conc.:20 µg/mL (IS/SS 20 µg/mL)
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Topaz 4 mm single taper w/wool (cat.# 23303)
Inj. Temp.:275 °C
Oven
Oven Temp.:45 °C (hold 0.5 min) to 285 °C at 56.5 °C/min to 305 °C at 6 °C/min to 330 °C at 60.5 °C/min (hold 2.5 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:330 °C
Quad Temp.:180 °C
Electron Energy:70 eV
Solvent Delay Time:1.3 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:39-550 amu
Scan Rate:9.8 scans/sec
InstrumentAgilent 7890B GC & 5977A MSD
NotesAnalyzed using a 208 V oven equipped with the GC Accelerator kit (cat.# 23849).