Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

MA DEP EPH Aromatics on Rxi-5Sil MS by Split Injection

GC_EV1487
PeakstR (min)Conc.
(µg/mL)
1.Naphthalene4.60125
2.2-Methylnaphthalene5.39125
3.2-Fluorobiphenyl (SS)5.82200
4.Acenaphthylene6.34125
5.2-Bromonaphthalene (SS)6.48200
6.Acenaphthene6.51125
7.Fluorene7.01125
8.Phenanthrene7.86125
9.Anthracene7.90125
10.o-Terphenyl (SS)8.19200
PeakstR (min)Conc.
(µg/mL)
11.Fluoranthene8.88125
12.Pyrene9.09125
13.Benz[a]anthracene10.57125
14.Chrysene10.63125
15.Benzo[b]fluoranthene12.55125
16.Benzo[k]fluoranthene12.61125
17.Benzo[a]pyrene13.23125
18.Indeno[1,2,3-cd]pyrene15.57125
19.Dibenz[a,h]anthracene15.64125
20.Benzo[ghi]perylene15.93125
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleMA EPH aromatic hydrocarbon standard (cat.# 31458)
MA fractionation surrogate spike mix (cat.# 31480)
o-Terphenyl (cat.# 31097)
Diluent:Methylene chloride
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Topaz 4 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:300 °C
Oven
Oven Temp.:55 °C (hold 0.5 min) to 150 °C at 20 °C/min to 250 °C at 30 °C/min to 280 °C at 5 °C/min to 345 °C at 30 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:4 mL/min
DetectorFID @ 340 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC