TNRCC 1006 Retention Time Marker Mix on Rxi-5Sil MS

GC_EV1492
PeakstR (min)Conc.
(µg/mL)
1.n-Hexane1.31200
2.n-Heptane2.09200
3.n-Octane3.32200
4.n-Decane5.46200
5.n-Dodecane7.11200
6.n-Hexadecane9.74200
7.n-Heneicosane12.33200
8.n-Octacosane15.14200
9.n-Pentatriacontane17.43200
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleTNRCC 1006 retention time marker mix (cat.# 31814)
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Topaz 4 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:300 °C
Oven
Oven Temp.:35 °C (hold 2 min) to 330 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:4 mL/min
DetectorFID @ 340 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890B GC