Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Melamine Spike on Rtx-5MS by GC/MS SIM

ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Samplemelamine, cyanuric acid, ammelide, ammeline, benzoguanamine
Conc.:10 µg/mL prederivatized
Injection
Inj. Vol.:1.0 µL splitless (hold 1 min)
Liner:3.5mm Gooseneck Splitless (cat.# 22286)
Inj. Temp.:280 °C
Oven
Oven Temp.:75 °C to 320 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:290 °C
Ionization Mode:EI