Melamine Spike on Rtx-5MS by GC/MS scan

ColumnRtx-5MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12623)
Standard/Samplemelamine, cyanuric acid, ammelide, ammeline, benzoguanamine
Conc.:10 µg/mL prederivatized
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:3.5mm Gooseneck Splitless (cat.# 22286)
Inj. Temp.:280 °C
Oven
Oven Temp.:75 °C (hold 1 min) to 320 °C at 15 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:290 °C
Ionization Mode:EI
Scan Range:50-450 amu