Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Pesticides in Dietary Supplements (Zoom of GCxGC-TOFMS Contour Plot)

GC_FF1188
ColumnRxi-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Rtx-200 1.5 m, 0.18 mm ID, 0.20 µm (cat.# 45001)
SampleMixed pesticide standard
Diluent:Toluene
Conc.:2 ng/µL
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Gooseneck Splitless (4mm) w/Wool (cat.# 22405)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:Rxi-5Sil MS: 80 °C (hold 1 min) to 310 °C at 4 °C/min (hold 1.5 min)
Rtx-200: 90 °C (hold 1 min) to 320 °C at 4 °C/min (hold 1.5 min)
Carrier GasHe, constant flow
Flow Rate:1.8 mL/min
Modulation
Modulator Temp. Offset:25 °C
Second Dimension Separation Time:4 sec
Hot Pulse Time:1.2 sec
Cool Time between Stages:0.8 sec
DetectorTOFMS
Transfer Line Temp.:290 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:4 min
Ionization Mode:EI
Acquisition Range:45 to 550 amu
Spectral Acquisition Rate:100 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesSee application note GNAN1338 for extraction and cleanup details. A 1.5 m length of the Rtx-200 column was trimmed from a 10 m column. Columns were connected with a Universal Press-Tight Connector (cat.# 20429). See chromatogram GC_FF1187 for full scale view.