Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Incurred Cypermethrins in a QuEChERS Extract of Marijuana by GCxGC on Rxi-5Sil MS and Rtx-200

GC_FF1203
PeaksRT 1 (sec.)RT 2 (sec.)
1.Cypermethrin 122921.50
2.Cypermethrin 223041.54
3.Cypermethrin 323101.53
4.Cypermethrin 423131.58
ColumnRxi-5Sil MS 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Rtx-200 1.3 m, 0.25 mm ID, 0.25 µm (cat.# 15020)
Standard/Sample
Diluent:Toluene
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Premium 4mm Single Taper w/Wool (cat.# 23303.1)
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:Rxi-5Sil MS: 80 °C (hold 1 min) to 310 °C at 5 °C/min
Rtx-200: 85 °C (hold 1 min) to 315 °C at 5 °C/min
Carrier GasHe, corrected constant flow (2 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:0.9 sec
Cool Time between Stages:0.6 sec
DetectorTOFMS
Transfer Line Temp.:290 °C
Analyzer Type:TOF
Source Temp.:225 °C
Electron Energy:70 eV
Mass Defect:-20 mu/100 u
Solvent Delay Time:5 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:100 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
NotesRtx-200 (cat.# 15020) is a 15 m column. A 1.3 m section was used as the second dimension column.