Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Palm Oil FAMEs on Stabilwax

GC_FF1266
PeakstR (min)Conc.
(wt.%)
Structural Nomenclature
1.Methyl laurate8.9633.0C12:0
2.Methyl myristate12.1062.0C14:0
3.Methyl palmitate15.34450.2C16:0
4.Methyl palmitoleate15.6681.5C16:1 (cis-9)
5.Methyl stearate18.2544.4C18:0
6.Methyl oleate18.62034.1C18:1 (cis-9)
7.Methyl linoleate19.2244.5C18:2 (cis9,12)
8.Methyl linolenate20.1180.1C18:3 (cis-9,12,15)
9.Methyl arachidate21.0010.2C20:0
ColumnStabilwax, 60 m, 0.25 mm ID, 0.25 µm (cat.# 10626)
SamplePalm oil standard (Supelco #46962)
Diluent:Hexane
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4 mm ID straight inlet liner w/ wool (cat.# 23300)
Inj. Temp.:225 °C
Oven
Oven Temp.:120 °C (hold 0.5 min) to 250 °C at 6 °C/min (hold 15 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
InstrumentAgilent 7890A GC
NotesThe sample was prepared using methanolic HCl.