Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Margarine on FAMEWAX

GC_FF1270
PeakstR (min)Structural Nomenclature
1.Methyl decanoate1.834C10:0
2.Methyl laurate2.751C12:0
3.Methyl myristate4.821C14:0
4.Methyl palmitate8.898C16:0
5.Methyl stearate14.687C18:0
6.Methyl oleate15.180C18:1 cis-9
7.Methyl elaidate15.285C18:1 trans-9
8.Methyl vaccenate15.384C18:1 cis-11
9.Methyl linoleate16.582C18:2 cis-9,12
10.Methyl linolenate18.603C18:3 cis-9,12,15
ColumnFAMEWAX, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12497)
SampleMargarine stick
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4 mm ID straight inlet liner w/wool (cat.# 23300)
Inj. Temp.:250 °C
Oven
Oven Temp.:160 °C to 240 °C at 2 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Inert
InstrumentAgilent 7890A GC & 5975C MSD
NotesThe sample was prepared using sodium methoxide.