Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Sunflower Oil on Rxi-65TG

GC_FF1294
PeakstR (min)
1.1,3-Palmitin-2-olein (POP)16.147
2.1-Palmitin-2-olein-3-stearin (POS)18.244
3.1,2-Olein-3-palmitin (POO)18.573
4.1-Palmitin-2-linolein-3-olein (PLO)19.047
5.1,2-Linolein-3-palmitin (PLL)19.561
6.1,2-Stearin-3-olein (SOS)20.709
7.Triolein (OOO)21.591
ColumnRxi-65TG, 30 m, 0.25 mm ID, 0.1 µm (cat.# 17108)
SampleSunflower oil
Conc.: ~50 mg of sunflower oil diluted to 10 mL with isooctane.
Injection
Inj. Vol.:1 µL split (split ratio 35:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:360 °C
Split Vent Flow Rate:35 mL/min
Oven
Oven Temp.:200 °C (hold 1 min) to 350 °C at 20 °C/min to 365 °C at 1 °C/min (hold 6 min)
DetectorFID @ 365 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890B GC