Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Rosemary Oil on Rtx-Wax

GC_FF1299
PeakstR (min)
1.α-Pinene1.473
2.Camphene1.544
3.β-Pinene1.617
4.β-Myrcene1.68
5.Limonene1.807
6.Eucalyptol1.844
PeakstR (min)
7.γ-Terpinene1.933
8.p-Cymene2.015
9.Camphor3.158
10.Caryophyllene3.637
11.α-Terpineol4.098
12.Borneol4.148
ColumnRtx-Wax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 12424)
Standard/SampleRosemary oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:230 °C
Oven
Oven Temp.:100 °C (hold 0.5 min) to 250 °C at 16 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesAll peaks were identified using the NIST MS EI spectra library (2005).