Lavender Oil on Rtx-Wax

GC_FF1301
PeakstR (min)
1.α-Pinene1.471
2.Camphene1.542
3.β-Pinene1.616
4.β-Myrcene1.679
5.D-Limonene1.805
6.Eucalyptol1.840
7.β-Ocimene1.857
8.α-Ocimene1.916
9.γ-Terpinene1.933
10.p-Cymene2.014
11.Terpinolene2.061
12.Octenyl acetate2.364
13.Hexyl butyrate2.554
PeakstR (min)
14.Camphor3.154
15.Linalool3.188
16.Linalyl acetate3.303
17.Geranyl butyrate3.559
18.4-Carvomenthenol3.575
19.Caryophyllene3.636
20.Lavandulol3.964
21.α-Terpineol4.098
22.Borneol4.147
23.β-Cubebene4.286
24.Geranyl acetate4.487
25.Lavandulyl acetate4.512
26.Geraniol4.994
ColumnRtx-Wax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 12424)
SampleLavender oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:230 °C
Oven
Oven Temp.:100 °C (hold 0.5 min) to 250 °C at 16 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:2 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:52 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
NotesAll peaks were identified using the NIST MS EI spectra library (2005).