Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Lemon Oil on Rxi-5Sil MS (20 m, 0.18 mm ID, 0.18 µm)

GC_FF1322
PeakstR (min)
1.α-Thujene 1.208
2.α-Pinene1.253
3.Camphene1.311
4.β-Myrcene1.363
5.β-Pinene1.402
6.Octanal1.412
7.α-Phellandrene1.477
8.α-Terpinene1.518
9.D-Limonene1.591
10.γ-Terpinene1.697
11.cis-β-Terpineol1.746
12.Terpinolene1.819
PeakstR (min)
13.Linalool1.85
14.Nonanal1.87
15.Citronellal1.901
16.α-Terpineol2.42
17.Z-Citral2.659
18.Geraniol2.699
19.E-Citral2.837
20.Neryl acetate3.413
21.Geranyl acetate3.543
22.β-Caryophyllene3.942
23.α-Bergamotene3.984
24.β-Bisabolene4.477
ColumnRxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602)
SampleLemon oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C (hold 0.25 min) to 320 °C at 17.5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).