Grapefruit Oil on Rxi-5Sil MS (20 m, 0.18 mm ID, 0.18 µm)

GC_FF1327
PeakstR (min)
1.α-Pinene1.248
2.β-Phellandrene1.362
3.β-Myrcene1.383
4.Octanal1.424
5.α-Phellandrene1.476
6.D-Limonene1.601
7.γ-Terpinene1.683
8.Linalool1.849
9.Decanal2.430
10.Z-Citral2.647
PeakstR (min)
11.Carvone2.720
12.E-Citral2.834
13.Geranyl acetate3.538
14.Copaene3.621
15.β-Cubebene3.694
16.Caryophyllene3.943
17.α-Caryophyllene4.181
18.Germacrene D4.347
19.Bicyclogermacrene4.440
20.δ-Cadinene4.564
ColumnRxi-5Sil MS, 20 m, 0.18 mm ID, 0.18 µm (cat.# 43602)
SampleGrapefruit oil
Diluent:Acetone
Conc.:5%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C (hold 0.25 min) to 320 °C at 17.5 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-5005
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).