Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Tea Tree Oil on Rxi-5Sil MS (10 m, 0.15 mm ID, 0.15 µm)

GC_FF1339
PeakstR (min)
1.α-Pinene0.514
2.Terpinolene0.612
3.p-Cymene0.622
4.D-Limonene0.625
5.Eucalyptol0.640
6.γ-Terpinene0.674
7.Pinocarveol0.840
8.Terpinen-4-ol0.926
PeakstR (min)
9.α-Terpineol0.945
10.Nerol0.991
11.Geraniol1.043
12.γ-Elemene1.274
13.Caryophyllene1.513
14.Aromadendrene1.562
15.α-Caryophyllene1.604
16.Alloaromadendrene1.617
17.Ledene1.691
ColumnRxi-5Sil MS, 10 m, 0.15 mm ID, 0.15 µm (cat.# 43815)
SampleTea tree oil
Diluent:Acetone
Conc.:1%
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 45 °C/min to 320 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.01 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
11.0035-50011
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Source Type:Inert
Source Temp.:230 °C
Quad Temp.:150 °C
InstrumentAgilent 7890A GC & 5975C MSD
NotesAll peaks were identified using the NIST MS EI spectra library (2005).