Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Benz[a]anthracene, Triphenylene, and Chrysene on Rxi-PAH After Repeated Matrix Injections

GC_FS0543
ColumnRxi-PAH, 40 m, 0.18 mm ID, 0.07 µm (cat.# 49316)
SampleCustom PAH standard
Diluent:Toluene
Injection
Inj. Vol.:0.5 µL pulsed splitless
Liner:Premium 2 mm single taper w/wool (cat.# 23316.1)
Inj. Temp.:275 °C
Pulse Pressure:80 psi (551.6kPa)
Pulse Time:0.6 min
Purge Flow:40 mL/min
Oven
Oven Temp.:110 °C (hold 1 min) to 210 °C at 35 °C/min to 260 °C at 3 °C/min to 350 °C at 10 °C/min (hold 4.5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorFID @ 360 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
InstrumentAgilent 7890A GC
NotesThe experiment was carried out as follows:

- Oysters and paprika were extracted using QuEChERS methodology.
- dSPE cleanup was performed on oyster extracts.
- Cartridge cleanup was performed on paprika extracts.
- Oyster and paprika extracts were injected, interspersed with solvent standard, to monitor performance.

1 m was trimmed from the column and the liner was changed during the course of the experiment.