Permethrins Resolution Stability on Low-Pressure GC Column Kit (500 Injections, Spinach Matrix)

GC_FS0574
PeakstR (min)Conc.
(ng/mL)
Parent IonProduct IonCollision Energy
1.cis-Permethrin7.829018315312
2.trans-Permethrin7.869018315312
ColumnLow-pressure GC column kit (factory-coupled restrictor column [5 m x 0.18 mm ID] and Rtx-5ms analytical column [15 m, 0.53 mm ID, 1 µm plus 1 m integrated transfer line on the outlet end]) (cat.# 11800)
SampleQuEChERS performance standards kit (cat.# 31152)
Diluent:Acetonitrile
Conc.:9 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID single taper inlet liner w/ wool (cat.# 23447)
Inj. Temp.:250 °C
Oven
Oven Temp.:70 °C (hold 1 min) to 320 °C at 35 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorTSQ 8000
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:325 °C
Solvent Delay Time:2 min
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesThe spinach matrix was prepared from 10 g of homogenized spinach extracted with QuEChERS EN salts (cat.# 25849) and cleaned up with dSPE containing magnesium sulfate, PSA, C18-EC, and GCB (cat.# 26219). The matrix extract was then spiked with 30 μL of each of the QuEChERS performance mixes for a final concentration of 9 ppm, and the internal standard triphenyl phosphate (TPP) was added at a final concentration of 10 ppm.
Between the first and last run, 500 injections of spinach extract spiked with internal standard (TTP) were made.