Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

GC Multiresidue Pesticide Standard #4-ONP on Rxi-5ms by GC-MS

GC_FS0603
PeakstR (min)
1.Allidochlor6.83
2.Pebulate9.37
3.N-(2,4-Dimethylphenyl)formamide9.63
4.Propachlor11.51
5.Cycloate11.80
6.Diallate 1*12.60
7.Diallate 2*12.80
8.Clomazone (Command)13.36
9.Propyzamide13.73
10.Triallate14.28
11.Propanil14.82
12.Dimethachlor14.87
13.Acetochlor15.03
14.Alachlor15.26
PeakstR (min)
15.Propisochlor15.35
16.Linuron15.79
17.Metolachlor16.10
18.Diphenamid16.61
19.Metazachlor16.91
20.Flutolanil18.07
21.Pretilachlor18.29
22.Oxadiazon18.40
23.Norflurazon19.93
24.Methoxychlor21.25
25.Fenpropathrin21.30
26.Tebufenpyrad21.37
27.Pyridaben23.30
28.Fluquinconazole23.41
29.Prochloraz23.48
*Diallate isomers numbered according to elution order.
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleGC multiresidue pesticide standard #4-ONP (cat.# 32566)
Diluent:Toluene
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4.0 mm ID Precision inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1555-5507
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:325 °C
Electron Energy:70 eV
Solvent Delay Time:5 min
Ionization Mode:EI
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesReconstructed chromatogram from select ions.