Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Custom Cleaning Solvents Standard on Stabilwax-MS

ColumnStabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673)
SampleCustom standard
Diluent:Methanol
Conc.:200 ppm
Injection
Inj. Vol.:1.0 µL split (split ratio 300:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:200 °C
Oven
Oven Temp.:60 °C to 220 °C at 30 °C/min
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.330-200
Transfer Line Temp.:200 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:2.3 min
Tune Type:PFTBA
InstrumentAgilent 7890A GC & 5975C MSD