Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Aromatics on Rt-TCEP

ColumnRt-TCEP, 60 m, 0.25 mm ID, 0.40 µm (cat.# 10999)
Sample
Conc.:500 ppm (except ethylbenzene @ 1,000 ppm)
Injection
Inj. Vol.:1.0 µL split
Inj. Temp.:200 °C
Split Vent Flow Rate:46 mL/min
Oven
Oven Temp.:60 °C (hold 5 min) to 100 °C at 5 °C/min (hold 10 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec @ 80 °C
DetectorFID @ 200 °C
NotesFID sensitivity: 6.4 x 10-11 AFS