Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Detailed Hydrocarbons Analysis Rtx-DHA-100

ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
using Rtx-5 DHA tuning column, 2.62m, 0.25mm ID, 1.0μm (cat.# 10165)
SampleCustom detailed hydrocarbons analysis (DHA) mix
Conc.: Neat
Injection
Inj. Vol.:0.1 µL split (split ratio 150:1)
Liner:Cup splitter with Siltek deactivation
Inj. Temp.:200 °C
Oven
Oven Temp.:5 °C (hold 15 min) to 50 °C at 5 °C/min (hold 50 min) to 200 °C at 8 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:2.3 mL/min
Linear Velocity:28 cm/sec
DetectorFID @ 250 °C
Notescatalog number 20709-214.1 Cup splitter with Siltek deactivation obsolete, recommend catalog number 20709