Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

m- and p-Xylene on Rtx-DHA-150

GC_PC1148
ColumnRtx-DHA-150, 150 m, 0.25 mm ID, 1.0 µm (cat.# 10149)
Samplem- and p-Xylene
Diluent:Methylene chloride
Injection
Inj. Vol.:1.0 µL split (split ratio 50:1)
Liner:4 mm ID cyclo inlet liner (cat.# 20706)
Inj. Temp.:275 °C
Oven
Oven Temp.:35 °C (hold 60 min) to 200 °C at 5 °C/min (hold 10 min)
Carrier GasH2, constant flow
Linear Velocity:44 cm/sec
DetectorFID @ 275 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC