Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

C10-C40 Alkanes on Rxi-5HT

GC_PC1159
Peaks
1.C10
2.C12
3.C14
4.C16
5.C18
6.C20
7.C22
8.C24
Peaks
9.C26
10.C28
11.C30
12.C32
13.C34
14.C36
15.C38
16.C40
ColumnRxi-5HT, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13924)
using Rxi guard column 2 m, 0.53 mm ID (cat.# 10054)
SampleSystem performance test standard mixture of n-alkanes (cat.# 31678)
Diluent:Hexane
Conc.:50 µg/mL
Injection
Inj. Vol.:1 µL cold on-column
Temp. Program:53 °C to 300 °C at 10 °C/min (hold 20 min)
Oven
Oven Temp.:50 °C to 300 °C at 10 °C/min (hold 20 min)
Carrier GasH2, constant flow
Linear Velocity:40 cm/sec @ 50 °C
Dead Time:1.25 min @ 50 °C
DetectorFID @ 330 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC