n-alkanes (C10-C40) on Rtx-Mineral Oil

GC_PC1181
Peaks
1.C10
2.C12
3.C14
4.C16
5.C18
6.C20
7.C22
8.C24
Peaks
9.C26
10.C28
11.C30
12.C32
13.C34
14.C36
15.C38
16.C40
ColumnRtx-Mineral Oil, 15 m, 0.32 mm ID, 0.10 µm (cat.# 18079)
using IP deactivated guard column 2 m, 0.32 mm ID (cat.# 10047)
SampleSystem performance test standard mixture of n-alkanes (cat.# 31678)
Diluent:Hexane/methylene chloride
Conc.:20 µg/mL each
Injection
Inj. Vol.:1.0 µL Direct
Inj. Temp.:58 °C
Oven
Oven Temp.:55 °C (hold 1.0 min) to 380 °C at 100 °C/min (hold 6.0 min)
Carrier GasH2, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 380 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesDirect cool on-column injection, programmed by oven tracking.