Gasoline Containing Ethanol on D3606 Application Column Set by ASTM D3606-10 (Modified)

GC_PC1224
ColumnD3606 application column (2 column set). Column 1: 6' (1.8 m), 1/8" OD, 2.0 mm ID, nonpolar Rtx®-1; Column 2: 16' (4.9 m), 1/8" OD, 2.0 mm ID, proprietary packing material (cat.# 83606-800)
SampleEthanol-containing gasoline with internal standard (IS)
Diluent:
Injection
Inj. Vol.:1.5 µL Direct
Inj. Temp.:150 °C
Oven
Oven Temp.:135 °C (hold 12 min)
Carrier GasHe, constant flow
Flow Rate:20.0 mL/min
DetectorTCD @ 200 °C
Notes2.38 minute backflush (must be determined for each GC system).