Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Separation of Ethanol and Aromatics from Paraffins in Gasoline with GCxGC on Rtx-DHA-150 and Stabilwax

GC_PC1227
Peaks1st Dim. RT (sec.)2nd Dim. RT (sec.)
1.Ethanol11101.84
2.Benzene23191.56
3.Toluene33301.56
ColumnRtx-DHA-150 150 m, 0.25 mm ID, 1.00 µm (cat.# 10149)
Stabilwax 1 m, 0.18 mm ID, 0.18 µm (cat.# 15118)
SampleGasoline
Conc.: "Neat" (undiluted)
Injection
Inj. Vol.:0.1 µL split (split ratio 200:1)
Liner:4 mm split Precision liner w/semivolatiles wool (cat.# 21022-231.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:Rtx-DHA-150: 40 °C (hold 1 min) to 245 °C at 1 °C/min
Stabilwax: 45 °C (hold 1 min) to 250 °C at 1 °C/min
Carrier GasHe, corrected constant flow (1.8 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:3 sec
Hot Pulse Time:0.8 sec
Cool Time between Stages:0.7 sec
DetectorFID @ 270 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Data Rate:200 Hz
InstrumentLECO GCxGC-FID
NotesThe Stabilwax column (cat.# 15118) is a 2 m column. A 1 m section was used as a second dimension column.

IP deactivated guard column (cat.# 10102) is a 1 m column. A 0.2 m section was used as a transfer line from the second dimension column to the detector.