Impurities in Methyl-t-Butyl Ether (MTBE) on Rtx-DHA-100 (ASTM D5441)

ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
SampleASTM D5441 MTBE contaminants mix
Conc.:1 % (w/w) each component in MTBE
Injection
Inj. Vol.:0.2 µL split (split ratio 200:1)
Liner:Cup splitter (cat.# 20710)
Inj. Temp.:200 °C
Oven
Oven Temp.:50 °C (hold 13 min) to 180 °C at 10 °C/min (hold 7 min)
Carrier GasHe, constant pressure (39.3 psi, 271.0 kPa)
Linear Velocity:23 cm/sec @ 50 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:24 mL/min
Make-up Gas Type:N2
Hydrogen flow:30 mL/min
Air flow:358 mL/min
InstrumentAgilent 7890A GC
NotesResolution between trans-2-pentene/tert-butanol and tert-butanol/cis-2-pentene exceeds the ASTM Method D5441 requirement of R ≥ 1.3.