Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Impurities in Methyl-t-Butyl Ether (MTBE) on Rtx-DHA-150 (ASTM D5441)

ColumnRtx-DHA-150, 150 m, 0.25 mm ID, 1.00 µm (cat.# 10149)
SampleASTM D5441 MTBE contaminants mix
Conc.:1 % (w/w) each component in MTBE
Injection
Inj. Vol.:0.2 µL split (split ratio 200:1)
Liner:Cup splitter (cat.# 20710)
Inj. Temp.:200 °C
Oven
Oven Temp.:60 °C (hold 13 min) to 180 °C at 10 °C/min (hold 20 min)
Carrier GasHe, constant flow
Flow Rate:2.2 mL/min
Linear Velocity:23 cm/sec @ 60 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:24 mL/min
Make-up Gas Type:H2
Hydrogen flow:30 mL/min
Air flow:358 mL/min
InstrumentAgilent 7890B GC
NotesResolution between trans-2-pentene/tert-butanol and tert-butanol/cis-2-pentene meets the ASTM Method D5441 requirement of R ≥ 1.3.