Sulfur Compounds on Rtx-1

ColumnRtx-1, 30 m, 0.32 mm ID, 4.00 µm (cat.# 10198)
Standard/SampleASTM D5623 standards 1 & 2 from DCG Partnership 1, Ltd.
Injection
Inj. Vol.:0.2 µL split (split ratio 10:1)
Liner:4 mm ID cup inlet liner with Siltek deactivation
Inj. Temp.:275 °C
Split Vent Flow Rate:24 mL/min
Oven
Oven Temp.:30 °C (hold 2 min) to 200 °C at 10 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:2.4 mL/min
Linear Velocity:38 cm/sec @ 30 °C
DetectorDP SCD @ 800 °C
Burner Vacuum:363 torr
Hydrogen Flow Rate:40 mL/min
Air Flow Rate:60 mL/min
InstrumentAgilent 7890B GC
NotesSolvent standard of sulfur components found in gasoline samples utilizing ASTM D5623.