Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Florida TRPH Standard Mix on Rxi-1HT

GC_PC1272
PeakstR (min)
1.n-Octane1.24
2.n-Decane2.71
3.n-Dodecane4.19
4.n-Tetradecane5.49
5.n-Hexadecane6.64
6.n-Octadecane7.68
7.n-Eicosane8.63
8.n-Docosane9.50
PeakstR (min)
9.n-Tetracosane10.30
10.n-Hexacosane11.04
11.n-Octacosane11.73
12.n-Triacontane12.38
13.n-Dotriacontane12.98
14.n-Tetratriacontane13.55
15.n-Hexatriacontane14.09
16.n-Octatriacontane14.60
17.n-Tetracontane15.08
ColumnRxi-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950)
SampleFlorida TRPH standard (cat.# 31266)
Diluent:Hexane
Conc.:50 ppm (1.25 ng on-column)
Injection
Inj. Vol.:1.0 µL split (split ratio 40:1)
Liner:Premium 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:40 °C (hold 1.0 min) to 400 °C at 20 °C/min (hold 10 min)
Carrier GasHe, constant flow
Linear Velocity:55 cm/sec
DetectorFID @ 400 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC