Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Simulated Distillation with Helium Carrier Gas on MXT-1HT SimDist by ASTM D2887-B (Accelerated)

GC_PC1303
ColumnMXT-1HT SimDist, 10 m, 0.53 mm ID, 0.88 µm (cat.# 70134)
SampleASTM D2887-12 calibration standard (cat.# 31674)
Injection
Inj. Vol.:0.02 µL cool on-column
Temp. Program:100 °C to 360 °C at 35 °C/min
Oven
Oven Temp.:60 °C to 360 °C at 35 °C/min
Carrier GasHe, constant flow
Flow Rate:26 mL/min
DetectorFID @ 360 °C
Make-up Gas Flow Rate:20 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:360 mL/min
InstrumentAgilent 7890B GC
NotesAccelerated analysis based on ASTM Method D2887 (Procedure B)