Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | Conc. (wt.%) | |
---|---|---|
1. | n-Heptane | 1.0 |
2. | Toluene | 0.5 |
3. | 2,3,3-Trimethylpentane | 0.8 |
4. | 2-Methylheptane | 1.0 |
5. | 4-Methylheptane | 1.0 |
6. | n-Octane | 1.0 |
Column | Rtx-DHA-50, 50 m, 0.20 mm ID, 0.50 µm (cat.# 10147) |
---|---|
Standard/Sample | D5134 column evaluation mixture (DCG Partnership I, LTD) |
Diluent: | 2-Methylpentane |
Injection | |
Inj. Vol.: | 0.2 µL split (split ratio 200:1) |
Liner: | Premium 4.0 mm ID Precision liner w/wool (cat.# 23305.5) |
Inj. Temp.: | 200 °C |
Oven | |
Oven Temp.: | 35 °C (hold 60 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 23 cm/sec @ 35 °C |
Detector | FID @ 250 °C |
---|---|
Make-up Gas Flow Rate: | 30 mL/min |
Make-up Gas Type: | N2 |
Hydrogen flow: | 40 mL/min |
Air flow: | 370 mL/min |
Instrument | Agilent 7890B GC |
Notes | The Rtx-DHA 50 column meets all the ASTM D5134-13 method requirements for efficiency, resolution, and polarity. Resolution (2-methylheptane/4-methylheptane) = 1.40 (method requirement ≥1.35) Efficiency (n-C8) = 239,631 plates (method specification ≥ 225,000) Column polarity = 0.5 (method 0.4±0.4) |