Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Simulated Distillation of Gasoline on MXT-1 (ASTM D7096-16)

GC_PC1328
PeakstR (min)Conc.
(vol.%)
1.n-Propane0.353-
2.2-Methylpropane0.419-
3.n-Butane0.475-
4.2-Methylbutane0.6766.5
5.n-Pentane0.7827.8
6.2-Methylpentane1.2147.5
7.n-Hexane1.4697.6
8.2,4-Dimethylpentane1.7427.6
9.n-Heptane2.4358.4
10.Toluene3.0179.5
PeakstR (min)Conc.
(vol.%)
11.n-Octane3.4386.7
12.p-Xylene4.0348.2
13.n-Propylbenzene4.8024.5
14.n-Decane5.2543.7
15.n-Butylbenzene5.6723.8
16.n-Dodecane6.8123.8
17.n-Tridecane7.5142.9
18.n-Tetradecane8.1732.9
19.n-Pentadecane8.7962.7
20.n-Hexadecane9.3832.8
21.n-Heptadecane9.9402.9
ColumnMXT-1, 15 m, 0.53 mm ID, 5.00 µm (cat.# 70177)
Injection
Inj. Vol.:0.5 µL split (split ratio 50:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:40 °C (hold 1 min) to 230 °C at 20 °C/min (hold 4 min)
Carrier GasHe, constant flow
Flow Rate:15 mL/min
DetectorFID @ 250 °C
Constant Column + Constant Make-up:30 mL/min
Make-up Gas Type:He
Hydrogen flow:40 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890B GC