Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Permanent Gases and Nitrous Oxide on Rt-Msieve 5A

GC_PC1346
PeakstR (min)
1.Oxygen1.25
2.Nitrogen1.32
3.Methane1.47
4.Carbon monoxide1.52
5.Nitrous oxide6.01
ColumnRt-Msieve 5A, 30 m, 0.53 mm ID, 50 µm (cat.# 19723)
Sample
Conc.: Varied (0.1-0.5% by volume)
Injection
Inj. Vol.:150 µL split (split ratio 10:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:200 °C
Oven
Oven Temp.:200 °C (hold 7 min)
Carrier GasHe, constant flow
Flow Rate:4 mL/min
DetectorTCD @ 250 °C
InstrumentAgilent 7890B GC