Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Permanent Gases on Rt-Msieve 5A (15 m x 0.53 mm x 50 μm)

GC_PC1347
PeakstR (min)Conc. (Mol%)
1.Helium0.63Balance
2.Argon0.943
3.Oxygen0.972
4.Nitrogen1.845
5.Methane2.464
6.Carbon monoxide8.115
ColumnRt-Msieve 5A, 15 m, 0.53 mm ID, 50 µm (cat.# 19721)
SamplePermanent gases (custom mix)
Injection
Inj. Vol.:30 µL split (split ratio 10:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:30 °C (hold 10 min)
Carrier GasH2, constant pressure (1.7 psi, 11.7 kPa)
DetectorTCD @ 200 °C
InstrumentAgilent 7890B GC
NotesResolution (Ar/O2) = 1.4
Asymmetry (CO) = 1.1

Note that the sample was introduced onto the column using syringe injection and a high split ratio, which results in a very narrow sample band. Any dead volume in the system (e.g., from using a sampling valve, large sampling loops, or even from using splitless injection) will result in peak broadening, which might decrease resolution between the peaks or even result in peak coelution.