Selectivity of Rt-BOND Porous Polymer PLOT Columns (C1-C2 Hydrocarbons)

GC_PC1360
PeaksConc. Mol%
1.Methane1.0
2.Ethylene1.0
3.Acetylene1.0
4.Ethane1.0
ColumnSee notes
SampleC1-C2 Hydrocarbons
Diluent:Helium (balance gas)
Injection
Inj. Vol.:50 µL split (split ratio 10:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:200 °C
Oven
Oven Temp.:40 °C (hold 3 min) to 100 °C at 20 °C/min
Carrier GasHe, constant flow
Flow Rate:4.5 mL/min
DetectorFID @ 200 °C
InstrumentAgilent 7890B GC
NotesColumns used:
Rt-Q-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19742)
Rt-QS-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19738)
Rt-S-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19746)
Rt-U-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19750)