Noble and Permanent Gases on Rt-Msieve 5A

GC_PC1362
Peaks
1.Helium
2.Neon
3.Hydrogen
4.Sulfur hexafluoride
5.Argon
Peaks
6.Oxygen
7.Nitrogen
8.Krypton
9.Methane
10.Carbon monoxide
11.Xenon
ColumnRt-Msieve 5A, 30 m, 0.53 mm ID, 50 µm (cat.# 19723)
SampleNoble and permanent gas standard
Conc.:2-10 mol%
Injection
Inj. Vol.:200 µL split (split ratio 10:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:35 °C (hold 1 min) to 150 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:5 mL/min
DetectorTCD @ 250 °C
InstrumentAgilent 7890B GC
NotesChromatogram A Analysis Conditions:
Inj. Vol.: 50 µL split (split ratio 50:1)
Conc.: 33 mol%
Inj. Temp.: 150 °C
Oven Temp.: 35 °C isothermal
Carrier Gas: N2, constant flow
Flow Rate: 3 mL/min
Detector: TCD @ 250 °C