Solvents on Rt-U-BOND

GC_PC1370
PeakstR (min)
1.Methanol3.25
2.Acetaldehyde3.79
3.Ethanol (solvent)-
4.Acetone7.31
5.2-Propanol7.89
6.Propanol8.79
7.Cyclohexane9.45
8.2-Methyl-2-propanol9.48
9.Benzene9.86
10.Butan-2-one9.94
11.2-Butanol10.56
PeakstR (min)
12.2-Methylpropan-1-ol10.99
13.Butanol11.51
14.2-Methyl-2-butanol12.07
15.2-Pentanol12.97
16.Acetal13.38
17.Pentanol13.95
18.4-Methylpentan-2-one14.06
19.Furfural14.89
20.4-Methyl-2-Pentanol15.02
21.3-Hexanol15.52
22.2-Hexanol15.80
23.Hexanol17.82
ColumnRt-U-BOND, 15 m, 0.32 mm ID, 10 µm (cat.# 19751)
Sample
Diluent:Ethanol
Conc.:4000 ppmv each analyte
Injection
Inj. Vol.:0.2 µL split (split ratio 50:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:60 °C (hold 0.5 min) to 190 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:4 mL/min
DetectorFID @ 200 °C
InstrumentAgilent 7890B GC