Peaks | |
---|---|
1. | Amphetamine |
2. | Methamphetamine |
3. | Nicotine |
4. | Cotinine |
5. | Caffeine |
6. | Benzphetamine |
Peaks | |
---|---|
7. | Ketamine |
8. | Phencyclidine |
9. | Methadone |
10. | Cocaine |
11. | Codeine |
12. | Scopolamine |
13. | Alprazolam |
Column | Rxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423) |
---|---|
Standard/Sample | 1000 μg/mL each |
Diluent: | Methanol |
Conc.: | 20 ng each compound on column |
Injection | |
Inj. Vol.: | 1.0 µL split (split ratio 50:1) |
Liner: | 4 mm ID single taper inlet liner, Siltek deactivation (cat.# 20799-214) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 220 °C at 15 °C/min to 330 °C at 10 °C/min (hold 5 min) |
Carrier Gas | He, constant pressure |
Linear Velocity: | 30 cm/sec |
Detector | Agilent 5973 MSD |
---|---|
Transfer Line Temp.: | 300 °C |
Analyzer Type: | Quadrupole |
Tune Type: | PFTBA |
Ionization Mode: | EI |
Scan Range: | 35-550 amu |