Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Residual Solvents (Class 2 - Mixture B) on Stabilwax (G16)

ColumnStabilwax, 30 m, 0.32 mm ID, 0.25 µm (cat.# 10624)
SampleResidual solvents class 2 - mix B (cat.# 36272)
in 20 mL headspace vial (cat.# 24685)
Diluent:Water
Injection
Inj. Vol.:1,000 µL headspace-syringe split (split ratio 5:1)
Liner:2 mm splitless (cat.# 20712)
Inj. Temp.:140 °C
Headspace-Syringe
Instrument:Overbrook Scientific HT200H
Syringe Temp.:100 °C
Sample Temp.:80 °C
Sample Equil. Time:45 min
Inj. Speed:8
Inj. Dwell: 5 sec.
Oven
Oven Temp.:50 °C (hold 20 min) to 165 °C at 6 °C/min (hold 20 min)
Carrier GasHe, constant flow
Flow Rate:2.15 mL/min
Linear Velocity:35.2 cm/sec
DetectorFID @ 250 °C