Calibration: Isopropylamine (5-100 ng on-column) on Rxi-624Sil MS

GC_PH1163
Peaks
1.Isopropylamine
ColumnRxi-624Sil MS, 30 m, 0.32 mm ID, 1.80 µm (cat.# 13870)
SampleIsopropylamine
Diluent:DMSO
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:5mm Single Gooseneck with Wool (cat.# 22973-200.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:50 °C (hold 1 min) to 200 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Linear Velocity:37 cm/sec
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesShown: 5 point calibration of isopropylamine; 5-100 ng on-column, r2 = 0.99996