Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

100 ng/mL Benzodiazepines in Hydrolyzed Urine on Ultra Biphenyl by LC-MS/MS

LC_CF0525
PeakstR (min)MRM1MRM2MRM3
1.7-Aminoclonazepam0.59286.0/250.3286.0/222.2286.0/121.0
2.7-Aminoflunitrazepam0.77284.2/135.2284.2/240.2284.2/226.1
3.Bromazepam1.10315.9/182.0315.9/209.1315.9/260.9
4.N-Desmethylflunitrazepam1.71300.0/254.1300.0/225.2300.0/198.0
5.Nitrazepam1.69282.1/236.1282.1/207.2282.1/180.0
6.Lorazepam1.71321.0/302.9321.0/275.1321.0/229.1
7.2-Hydroxyethylflurazepam1.79333.1/246.1333.1/166.0333.1/109.1
8.Nordiazepam1.78271.0/140.0271.0/208.0271.0/225.9
9.Estazolam1.88294.9/267.1294.9/241.0294.9/138.0
10.Desalkylflurazepam1.90289.0/140.0289.0/165.1289.0/226.0
PeakstR (min)MRM1MRM2MRM3
11.Clonazepam1.92315.9/270.2315.9/214.1315.9/207.1
12.Alprazolam1.99309.4/281.1309.4/205.0309.4/274.1
13.Midazolam1.26326.1/291.2326.1/244.2326.1/209.0
14.Triazolam2.06343.1/315.0343.1/308.2343.1/239.2
15.Temazepam2.13301.2/283.1301.2/255.1301.2/228.1
16.Flurazepam1.34388.1/315.1388.1/288.2388.1/317.1
17.Flunitrazepam2.24314.4/211.0314.4/239.1314.4/268.1
18.Clobazam2.26301.1/259.1301.1/224.0301.1/181.1
19.Diazepam-D5 (IS)2.42290.0/154.1----
20.Diazepam2.45284.8/193.1284.8/222.2284.8/257.1
ColumnUltra Biphenyl (cat.# 9109552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:40 °C
Sample
Diluent:Starting mobile phase + 30 ng/mL IS
Conc.:100 ng/mL sample was diluted 20x prior to injection
Inj. Vol.:30 µL
Mobile Phase
A:Water + 2 mM ammonium formate + 0.2% formic acid
B:Acetonitrile + 10% water + 2 mM ammonium formate + 0.2% formic acid
Time (min)Flow (mL/min)%A%B
0.000.67030
3.00.64060
3.10.6595
3.50.6595
3.60.67030
4.5stop
DetectorAB SCIEX API 4000 MS/MS
Model #:API 4000
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Ion Spray Voltage:2 kV
Curtain Gas:40 psi (275.8 kPa)
Gas 1:40 psi (275.8 kPa)
Gas 2:40 psi (275.8 kPa)
Source Temp.:600 °C
Mode:MRM
InstrumentApplied Biosystems/MDS Sciex LC-MS/MS System
NotesMS/MS instrument was operated in scheduled MRM mode.