Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.

Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Benzodiazepines on Raptor Biphenyl by LC-MS

LC_CF0563
PeakstR (min)SIM
1.Bromazepam1.23317.95
2.Flurazepam1.36388.10
3.Oxazepam2.05287.05
4.Nitrazepam2.16282.05
5.Lorazepam2.20321.00
6.Alprazolam2.55316.00
PeakstR (min)SIM
7.Clonazepam2.61309.05
8.Triazolam2.71343.05
9.Temazepam3.02301.05
10.Flunitrazepam3.25314.05
11.Clobazam3.28301.05
12.Diazepam3.54285.05
ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Temp.:40 °C
Sample
Diluent:Water:acetonitrile (70:30)
Conc.:5 µg/mL
Inj. Vol.:1 µL
Mobile Phase
A:2 mM Ammonium formate + 0.2% formic acid in water
B:0.2% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.57030
3.00.56040
3.50.55050
3.60.5595
4.00.5595
4.10.57030
5.00.57030
DetectorShimadzu LCMS-2010EV
Acquisition Type:Scan/SIM
Probe Voltage:1.5 kV
Nebulizing Gas Flow:1.5 L/min
Drying Gas Pressure:0.1 MPa
CDL Temp.:300 °C
Heat Block:300 °C
Interface:ESI+
InstrumentShimadzu UFLCXR