Coenzyme Q10 in Serum on Raptor Biphenyl by LC-MS/MS

LC_CF0611
PeakstR (min)Precursor IonProduct Ion
1.Ubiquinol-100.98882.82197.18
2.Ubiquinone-10-D91.92872.89206.16
3.Ubiquinone-101.95863.86197.12
ColumnRaptor Biphenyl (cat.# 9309552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 5 µm (cat.# 930950252)
Temp.:40 °C
Sample
Diluent:See sample preparation
Conc.: Endogenous levels in serum
Inj. Vol.:5 µL
Mobile Phase
A:0.1% formic acid, 5 mM ammonium formate in 50:50 water:acetonitrile
B:0.1% formic acid, 5 mM ammonium formate in methanol
Time (min)Flow (mL/min)%A%B
0.000.6595
2.500.60100
2.510.6595
4.000.6595
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation Procedure: