Welcome to the New Restek.com! Take a tour of the new site and feel free to send us your feedback.
Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Metanephrines on Raptor FluoroPhenyl by LC-MS/MS

LC_CF0642
PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Normetanephrine1.15166.1121.1134.0
2.Metanephrine1.52180.1165.1148.3
ColumnRaptor FluoroPhenyl (cat.# 9319A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:30 °C
Sample
Conc.: 20 ng/mL in water
Inj. Vol.:5 µL
Mobile Phase
A:Water + 0.2% formic acid
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.4982
2.000.46040
2.010.4982
6.000.4982
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC